4-4
WARNING
The infusion set should be firmly loaded into the slot, and not jutting on
the outside of the slot.
Before using this infusion pump, the infusion pump, infusion set and
other accessories should be loaded correctly.
4.2.3 Change the Infusion Set
Follow the steps below to change the infusion set:
1. To prevent animal injury due to free flow, before changing the infusion set or
extruded tube, please shut down the Robert clip (or flow rate regulator).
During infusion, press
to stop the pump.
2. Pull the door holder, open the door, pull the free flow clamp button upward left,
and take out the loaded infusion set.
3. Please refer to
4.2.2 Load the Infusion Set
to reload the infusion set.
4.2.4 Change the Infusion Bottle (Bag)
Follow the steps below to change the infusion bottle (bag):
1. To prevent animal injury due to free flow, before changing the infusion bottle
(bag), please shut down the Robert clip (or flow rate regulator). During
infusion, press
to stop the pump.
2. Take out the loaded infusion bottle (bag), and reload it.
4.2.5 Select Infusion Set Brand
On the [
Set selection
] screen, press
to select the infusion set brand,
and press
for confirmation.
CAUTION
Please confirm that the current selected brand is the same as the brand
actually used, or its accuracy cannot be guaranteed.
Summary of Contents for BeneFusion VP3 Vet
Page 1: ...BeneFusion VP3 Vet Infusion Pump Operator s Manual...
Page 2: ......
Page 73: ...A 11 Unit of Rate ml h...
Page 90: ...F 1 F Declaration of Conformity P N 046 013101 00 4 0...
Page 91: ......
Page 92: ......