B-2
Guidance and Declaration - Electromagnetic Emissions
The device is intended for use in the electromagnetic environment specified below. The
customer or the user of the device should assure that it is used in such an environment.
Emission test
Compliance Electromagnetic environment - guidance
Conducted and radiated
RF EMISSIONS CISPR 11
Group 1
The device uses RF energy only for its
internal function. Therefore, its RF
emissions are very low and are not likely to
cause any interference in nearby electronic
device.
Conducted and radiated
RF EMISSIONS CISPR 11
Class A
The device is suitable for use in all
establishments other than domestic and
those directly connected to the public
low-voltage power supply network that
supplies buildings used for domestic
purposes.
Harmonic distortion
IEC 61000-3-2
Class A
The device is suitable for use in all
establishments, including domestic
establishments and those directly
connected to the public low-voltage power
supply network that supplies buildings used
for domestic purposes.
Voltage fluctuations and
flicker IEC 61000-3-3
Complies
NOTE
The device needs special precautions regarding EMC and needs to be
installed and put into service according to the EMC information provided
below.
Other devices may affect this device even though they meet the
requirements of CISPR.
The EMISSIONS characteristics of this device make it suitable for use in
industrial areas and hospitals (CISPR 11 class A). If it is used in a
residential environment (for which CISPR 11 class B is normally
required) this device might not offer adequate protection to
radio-frequency communication services. The user might need to take
mitigation measures, such as relocating or re-orienting the device.
If the essential performance is lost or degraded, it may be necessary to
take mitigation measures, such as re-orienting or relocating the ME
EQUIPMENT or ME SYSTEM or shielding the location or stopping using
the infusion pump system and contact the service personnel.