Battery
WARNING
The battery can not be disassembled. The battery should be changed by
maintenance staff designated by the company only. Changing the
battery incorrectly or changing battery by personnel who has not
received suitable training may cause such danger as overtemperature,
fire or explode.
Use only manufacturer-specified battery for this device. Use of a
different battery may cause such danger as overtemperature, fire or
explode.
NOTE
When the built-in battery and external power encounter failure, the
display will off, a high level alarm will be triggered, the buzzer will sound
out and the red alarm light will continuously flash.
The infusion pump is configured with rechargeable Lithium batteries to ensure that
the infusion pump can be used normally under the condition of the animal's
migration within the hospital or during the circumstance of a power failure. When the
infusion pump switches to the AC power, the battery can be charged regardless of
whether the infusion pump is on or off. The battery is chargeable only within the
infusion pump. During charge, the battery icon in the upper-right corner of the
screen floats up and down. If the battery icon stops floating and is completely filled,
it indicates that the battery is fully charged. Under the condition of a sudden power
failure, the pump will automatically use the battery to provide power as a backup.
The battery icon on the screen indicates the condition of the battery:
The battery jar of the infusion pump is installed with batteries, and
the white fill area indicates the quantity of electricity.
Low battery electric quantity indicates that charging is needed.
When the battery is empty, charging is needed immediately.