Page 74
AiRanger SPL
PL-573
Enhancement Pa
ra
m.
P818 (V)
Profile Pointer Distance
Observe the distance (in P005 Units) between the transducer face and the Profile
Pointer.
Values:
0.000 to 9999
P819 (V)
Profile Pointer Amplitude
Observe the amplitude (in dB above 1
µ
V rms) of the Echo Profile at the Profile
Pointer position.
Values:
0 to 99
P820
Algorithm
Use this feature to select the
Algorithm(s)
(mathematical operations) which the Sonic
Intelligence™ echo selection is to be based on.
This value is automatically altered by Material (P002).
If the wrong echo is processed, select an alternate algorithm, while observing the
resultant echo processing displays.
To select an Algorithm...
1. Press
to display the Auxiliary Function symbol,
2. Press
or
to display the desired Reading display symbol(s),
3. Press
(when the desired Algorithm is displayed)
The material/vessel type most likely to benefit from a particular algorithm is included
with the following option values.
Value
Description
ALF
long range Area, Largest, and First average (general purpose, solids)
A
long range Area only (coarse, heaped solids)
L
long range Largest only (liquids, open vessels)
F
long range First only (liquids, closed vessel)
AL
long range Area and Largest average (fine heaped solids)
AF
long range Area and First average (coarse flat solids)
LF
long range Largest and First average (general purpose, liquids)
bLF
short range Largest or First (general purpose)
bL
short range Largest only (solids and open vessel liquids)
bF
short range First only (closed vessel liquids)
Summary of Contents for AiRanger SPL PL-573
Page 1: ...AIRANGER SPL Instruction Manual PL 573 March 2001 IRANGER SPL 33455730 Rev 1 2 ...
Page 8: ...Page 8 AiRanger SPL PL 573 Specifications ...
Page 24: ...Page 24 AiRanger SPL PL 573 Installation ...
Page 86: ...Page 86 AiRanger SPL PL 573 Enhancement Param ...
Page 102: ...Page 102 AiRanger SPL PL 573 Maintenance ...
Page 108: ...Page 108 AiRanger SPL PL 573 Troubleshooting ...