Chapter 7 Analysis System
123
Positive overshoot
Positive overshoot =[(max - high) / amplitude]*100%
Negative overshoot
Negative overshoot =[(low - min) / amplitude]*100%
Phase
Timing measurement. The amount of time that one waveform leads or lags another waveform, expressed in degrees
where 360°comprises one waveform cycle.
Peak-peak
In the entire waveform measurement, peak-peak = max - min
Amplitude
In the entire waveform measurement, amplitude = high (100%) - low (0%)
The figure below shows voltage measurement points.
The channel probe type setting is used to set the measurement unit for each input channel to Volts or Amperes. Refer
to “
5.4.4 Set Probe Type
”.
High
Take 100% in the entire waveform, and calculated using either the min/max or histogram method.
Low
Take 0% in the entire waveform, and calculated using either the min/max or histogram method.
Max
Highest positive peak measured over the entire waveform
Min
Highest negative peak measured over the entire waveform
RMS
True root mean square value over the entire waveform
C RMS
True root mean square value of the first cycle in the waveform
Mean
Arithmetic mean over the entire waveform
C mean
Arithmetic mean over the first cycle in the waveform
Note: