U
SING
F
ILLER
R
ODS
1. Clean the sample tube, balance tube, and filler rods; label your sample and balance tubes.
2. Prepare the sample and place it into the sample tube; insert the shorter filler rod into the
sample tube.
If using bulbous sample tubes, use the same size filler rod as used in the balance tube.
3. Ensure that the filler rods are equidistant from the top of each tube. If they are not, add or
remove sample until they are the same distance from the top of the tube.
Packing of some powders may restrict gas access to the powder and cause slower
equilibration and/or lower results.
4. Attach the sample tube to the analysis port and the balance tube to the balance port.
5. Prepare the analysis Dewar and place it on the elevator.
6. Close the sample compartment door and start the analysis.
A Analyze Samples with a Total Surface Area of
1.0m2 or Less
Gemini VII Operator Manual
239-42828-01 (Rev H ) — May 2021
A - 3