1 A
NALYZER
C
OMPONENTS
Parts and accessories can be found online at
.
F
RONT
C
OMPONENTS
A. Dewar
B. Sample safety shield
C. Power indicator light
D. Balance port and balance tube
E. P
0
port and tube
F. Sample port and sample tube
G. Tube position diagram
H. Elevator
Components
Description
Dewar
For housing cryogens used for analysis. The Dewar rests on an
elevator.
Sample safety shield
Closes to cover the Dewar.
Power indicator light
Blinks when power is applied to the analyzer; illuminates when
the analysis program is initiated and ready for operation.
Balance port and balance
tube
The balance tube is identical to the sample tube. This design
negates free-space errors introduced by thermal gradient vari-
ations or by initial mis-measurement of free space.
P
0
port and tube
Models 2390p and 2390t only. For measuring the saturation
pressure.
Front Components
1 Analyzer Components
Gemini VII Operator Manual
239-42828-01 (Rev H ) — May 2021
1 - 1