10 Troubleshooting
8. Click the
Gas Pressure Test - 3
tab to display its report.
On the report, the
Minimum field
should read
Passed
indicating all valves are in the proper
state for operation.
If the
Minimum
field indicates
Failed
, one or more valves is not in the proper position. Set
the valves as shown at the beginning of this section and ensure the appropriate pressure is
displayed on the low pressure gauge.
If running the test again, close the gas cylinder valve before starting the test.
9. Click
Close
to close the test report. Click
Close
again to close the test.
10 - 10
Gemini VII Operator Manual
239-42828-01 (Rev H ) — May 2021