S
AMPLE
P
ORT
F
RIT
R
EPLACEMENT
The equipment images in this topic may differ slightly from your equipment; however,
the instructions are the same unless otherwise noted.
A frit is located in the connecting nut attached to each analysis port. If the frit becomes
contaminated, the contaminant may adsorb or desorb during analysis, affecting the results. A
contaminated frit on the analysis port may be indicated as a leak or a free space reading much
lower than normal.
Use the proper size frit for the sample port. The analyzer will not operate properly if an
incorrect size is used.
1. Go to
Unit [n] > Enable Manual Control
. Ensure a checkmark displays to the left of the
menu item. If the analyzer schematic does not display, go to
Unit [n] > Show Instrument
Schematic.
2. Right click on the valve of the appropriate port. If the valve is open, click
Close
to close the
valve.
3. Use a wrench to remove the connecting nut from the sample port. Pry out the filter (frit) and
O-ring.
To avoid degassing problems, the frit should be clean and should not be touched with
bare hands.
Sample Port Frit Replacement
Gemini VII Operator Manual
239-42828-01 (Rev H ) — May 2021
10 - 23
Summary of Contents for GEMINI VII
Page 1: ...OPERATOR MANUAL 239 42828 01 May 2021 Rev H GEMINI VII SURFACE AREA ANALYZER ...
Page 10: ......
Page 16: ......
Page 92: ......
Page 126: ...6 About Reports SAMPLE LOG REPORT 6 34 Gemini VII Operator Manual 239 42828 01 Rev H May 2021 ...
Page 127: ...T PLOT REPORT t Plot Report Gemini VII Operator Manual 239 42828 01 Rev H May 2021 6 35 ...
Page 128: ...6 About Reports VALIDATION REPORT 6 36 Gemini VII Operator Manual 239 42828 01 Rev H May 2021 ...
Page 234: ......
Page 242: ......
Page 244: ......
Page 247: ......
Page 248: ......