e. Reconnect the gas line to the regulator.
f. Use two 7/16 in. (11 mm) wrenches to tighten the gas line connection. Hold one wrench
fitting steady and the other to tighten the connector nut.
6. Set the analyzer pressure by doing the following:
a. Turn the
Regulator Control
knob clockwise until the low pressure gauge indicates the
appropriate pressure. See the
Regulator Pressure Settings
table in
.
b. Open the regulator
Shut-off
valve.
c. Open the gas cylinder
Shut-off
valve and flow gas for 10 to 30 seconds.
d. Close the gas cylinder
Shut-off
valve.
e. Close the gas cylinder valve.
7. If the gas line to the instrument inlet was previously disconnected, reconnect it now.
Replace a Gas Cylinder
Gemini VII Operator Manual
239-42828-01 (Rev H ) — May 2021
10 - 13