M
ATCH
T
RANSDUCERS
Unit [n] > Calibration > Match Transducers
This process should not be performed when the main transducer reading is abnormal.
This option is disabled for the Gemini 2390a model.
Use to evacuate the system and zero the pressure transducers, then adjust the scale to match
them to the manifold transducer near full scale pressure.
A blank sample tube or small plug must be installed on each selected port prior to start-
ing this process.
1. Install a blank sample tube or small plug on each port.
2. Click
Start
. The window closes when the operation is complete. Click
OK
when the process
is complete.
Match Transducers
Gemini VII Operator Manual
239-42828-01 (Rev H ) — May 2021
9 - 3