
9 Calibration
B
ALANCE
S
ERVO
V
ALVE
Unit [n] > Calibration > Balance Servo
Use to calibrate the servo valve to the manifold pressure transducer. The servo valve should
always be recalibrated after a pressure calibration has been performed. The pressure transducer
should be calibrated before starting this calibration procedure.
Ensure the pressure transducer has been calibrated before performing this pro-
cedure. Go to
Unit [n] > Unit Configuration
and view the calibration information.
Contact your Micromeritics Service Representative if calibration dates are not listed.
Click
Start
. The window closes when the calibration is complete. Click
Cancel
to stop the
calibration process.
9 - 2
Gemini VII Operator Manual
239-42828-01 (Rev H ) — May 2021
Summary of Contents for GEMINI VII
Page 1: ...OPERATOR MANUAL 239 42828 01 May 2021 Rev H GEMINI VII SURFACE AREA ANALYZER ...
Page 10: ......
Page 16: ......
Page 92: ......
Page 126: ...6 About Reports SAMPLE LOG REPORT 6 34 Gemini VII Operator Manual 239 42828 01 Rev H May 2021 ...
Page 127: ...T PLOT REPORT t Plot Report Gemini VII Operator Manual 239 42828 01 Rev H May 2021 6 35 ...
Page 128: ...6 About Reports VALIDATION REPORT 6 36 Gemini VII Operator Manual 239 42828 01 Rev H May 2021 ...
Page 234: ......
Page 242: ......
Page 244: ......
Page 247: ......
Page 248: ......