XZR400 Series User’s Manual
Michell Instruments
21
INSTALLATION
2.9.4 D-Sub DA15 plug
This optional D-Sub DA15 male plug (15 pin) is used to connect a second 4-20 mA
output, a total pressure adjustment inlet, and the analyzer's auto adjustment terminals,
as well as automatic scale switching operations.
2.9.5 Sample
Path
The MSRS sensor is placed inside an oven in which the gases to be analyzed are
circulated. The oven consists of a gas inlet head and a outlet plate. 3 Viton O-rings
ensure the sealing of this device (2 for the inlet and 1 for the outlet).
The internal sample path consists of:
• 2
fl
ow control valves: sensor
fl
ow and by-pass
fl
ow
•
1 sealing head
•
1 outlet plate
•
1 oven tube
• 1
electronic
fl
ow meter
•
1 MSRS sensor
•
2 Swagelok 6mm stainless steel bulkhead unions (gases inlet and outlet)
- under the analyzer
XZR400A2
Sensor flow adjustment valve
O
2
sensor
By-pass flow adjustment valve
Barometric pressure sensor
6mm bulkhead union
6mm bulkhead union
Sampling gas inlet
Sampling gas outlet
Electronic flowmeter 0-12 l/h
Figure 21
XZR400A2 Sample Path