XZR400 Series User’s Manual
50
97472 Issue 3, August 2018
OPERATION
Misc Tab
The Misc tab of the System screen concerns the oven temperature setpoint, the menu
display language and the access code modi
fi
cation.
•
Touch the + and – area to set
Hour
System
5.5
634ºC
Setpoint
Language
New access code
English
****
Misc
.
Area
Function
Setpoint
Enter the oven temperature value that the analyzer will use to operate.
The default value is 634°C. An unsuitable temperature setting reduces
the performance of the analyzer and can damage the MSRS oxygen
measuring cell.
Language
Touch the
Language
area. In the opened window, change the
displayed language. The new language will be applied at the next
start.
NOTE: the language selection is proposed if this choice was
not selected at the analyzer start.
New
access
code
Modi
fi
es the expert level code. By default, the expert code is
0
.
Warning: If the access code is forgotten the equipment will
need to be returned to the factory for Expert level code reset.
Factory
settings
The analyzer parameter values defaults are set at the factory, namely
•
Oven temperature: 634°C
•
Alarm No. 1 High mode, threshold: 3 ppm
•
Alarm No. 2 High mode, threshold: 3.5 ppm
•
Low scale and high scale, 4-20 mA output No. 1: 0.1 and
10 ppm
•
Low scale and high scale, 4-20 mA output No. 2: 0.1 and
1000 ppm