XZR400 Series User’s Manual
Michell Instruments
19
INSTALLATION
2.9.1 Gas Sample Inlet and Outlet Fittings
The gas path consists of 2
fl
ow control valves (sensor
fl
ow and by-pass
fl
ow), a sensor
and an electronic
fl
owmeter. For the gas inlet and the outlet connections, Swagelok
stainless steel 6mm bulkhead unions are used, which guarantees perfect sealing and
easy removal.
The evacuation of the analyzed sample must be carried out at
atmospheric pressure, otherwise the total pressure adjustment
option is required.
The measurement of hazardous gas samples require the
analyzer's outlet to be outside the analysis area and must be in
accordance with local regulations.
Sensor flow
adjustment
Bypass flow
adjustment
Inlet for gas
to be
analyzed
Analyzed
gas
outlet
Sensor
Electronic
flow meter
Figure 19
Gas Circuit Diagram for XZR400A2