XZR400 Series User’s Manual
34
97472 Issue 3, August 2018
OPERATION
3.3 Warm-Up
Period
Upon power up, the screen appears as shown below. The analyzer performs a series of
internal checks for about 5 seconds.
NOTE: Touch the French or English area to display the menu in the required
language.
French
1
Software version V0.1.7.ver
English
Figure 29
Start-up Screen
Wait about 15 minutes until the oven temperature reaches 634°C, indicated by the oven
temperature line.
The low temperature alarm is displayed digitally throughout the rise in oven temperature;
the general alarm contact is activated.
NOTE: A fl ow greater than 3.5 l/h prevents the correct rise in oven temperature.
Oven
Flow
Low temperature alarm
48.8ºC
7.5 l/h
Figure 30
Oven Temperature Screen