XZR400 Series User’s Manual
46
97472 Issue 3, August 2018
OPERATION
Main Tab
Main
Level
Auto adjustment
5.4.1
External trigger
T0 adj
Adjustment period
Unit
No
12:00
1 d
d
Process
Enable
Start
Process
Area
Function
Enable/Disable
Enables/disables the launch of the automatic adjustment.
Start
Management of the start of the auto-adjust procedure.
•
Start
: the auto adjust sequence is launched immediately,
regardless of the other settings.
•
T0 start
: the auto-adjust sequence is launched at the time
speci
fi
ed by the area
T0 Adjust.
The interval is set by
Adjustment period
(see below).
External trigger
Management of an external launch control button (push-button or
dry contact from an external PLC) controlling the launch of the auto
adjust sequence (as opposed to the automatic launch of the Inter
Start and Inter Adjust commands).
•
No
: pressing the external button connected to terminals
5-6 of the D-Sub DA15 connector or the 8 pluggable
pin connector will not allow the launch of the auto adjust
sequence.
•
Yes
: pressing the external button connected to terminals
5-6 of the D-Sub DA15 connector or the 8 pluggable pin
connector will allow the launch of the auto adjust sequence.
T0 adj
Sets the start time of the next auto-adjust.
Adjustment
period
Adjustment interval
Sets the time interval in hours or days between two auto-adjust
sequences. See below.
Unit
d (day) or hr (hour)
De
fi
nes the time unit of the Adjustment interval.
Inter adjust:
Intervals in hours or days
between automatic
adjustments
Automatic
adjustment
sequence
Automatic
adjustment
sequence
To
next start
time