XZR400 Series User’s Manual
44
97472 Issue 3, August 2018
OPERATION
The Alarm 1 or Alarm 2 Tab
This screen is used to set the alarm function mode (high, low), the threshold and
hysteresis for the selected alarm 1 or 2.
It is used mainly to enable or disable the alarms during the adjustment phases and
during normal analyzer operation.
•
Touch the area to select it and change it.
•
Enter any new numeric value on the virtual keypad.
• Click
OK
to con
fi
rm or
Cancel
to discard the changes.
Main
Alarms
5.3
Low
Type
Threshold
Unit
Hysteresis
1.0%
%
0.1%
Alarm 2
Alarm 1
Figure 39
Alarm Screen (Alarm 1)
Area
Function
Type
Sets the behaviour for alarm (1 or 2 according to the highlighted tab).
•
Low
: the alarm will be activated when the measurement value
is equal or below the threshold set for the Alarm threshold
line. The alarm must be disabled when the measurement
value is above the threshold set for the Alarm threshold line.
•
High
: the alarm will be activated when the measurement value
is equal to or above the threshold set for the Alarm threshold
line. The alarm must be disabled when the measurement
value is below the threshold set for the Alarm threshold line.
Threshold
Con
fi
gures the alarm threshold value. Select the value and unit in relation
to the future measurement.
Hysteresis
Sets the alarm 'delay' value to return to alarm-off condition. The greater
this value is, the more the return to alarm-off condition value must be
delayed. For example, with a hysteresis of 3ppm over a 15ppm alarm
value, the alarm is triggered at 15ppm, but returns to the alarm-off
position at a 12ppm value (15ppm - 3ppm).
Returns to the Main Menu Screen.