XZR400 Series User’s Manual
52
97472 Issue 3, August 2018
OPERATION
3.6.6 COMM
3.6.6.1 RS232
Area
Function
Frame period
Sets the delay, in seconds, between two RS232 signal emissions, from
0 (no emission) to 999 seconds (one emission every about 16 minutes).
Returns to the Main Menu Screen.
3.6.6.2 RS485
Only available for analyzers equipped with this output.
Comm Screen
The COM RS485 screen sets the RS485 output parameters.
Comm
5.7
Speed
Data
Parity
Stop
Modbus address
1
9600 bps
8 bits
none
1
RS485
Figure 42
Comm Screen
Area
Function
RS485
Indications of the RS485 interface connection parameters that cannot
be modi
fi
ed.
•
Speed: 9600 bauds
•
Number of bits
• Parity:
None
• Stop:
1
ModBus
address
The analyzer's ModBus slave address. Enter a value between 1 and 255,
including terminals.