XZR400 Series User’s Manual
Michell Instruments
37
OPERATION
3.5
Control Parameters Display
On the Measurement Display Screen (3) touch the
VIEW
area
The Control Parameters Screen (3.1) will appear.
Adjustment
CO
Adjustment
Maintenance
3.1
634.0ºC
30.7ºC
-122.76 mV
1036 mBar
2.5 L/h
0.00
Oven
Room temp
Sensor Voltage
Atm pressure
Flow
Control parameters
O2 27.35 ppm
Figure 33
Control Parameters Screen
This displays the following information about the analyzer.
Oven
Current temperature of the hot joint of the thermocouple
corresponding to oven temperature. This must be 634°C.
Temperatures 30°C below the reference and 50°C above the
reference trigger a general alarm.
Room Temperature
Measure taken on the motherboard. This measurement is only
used to compensate for the cold solder joint of the thermocouple.
Sensor Voltage
Measurement in mV of the voltage supplied by the measuring
cell, which must be between -300 and +250 mV. A value greater
than the upper limit generates a MSRS fault alarm.
Atm pressure
The pressure at the analyzer outlet must be near atmospheric
pressure (between 800 and 1750 mbar). A value above 1200
mbar and under 800 mbar generates an oxygen measurement
error.
NOTE: If the pressure is greater than 1750 mbar,
the atmospheric pressure sensor can be damaged
irreversibly.
Flow
The sample's
fl
ow in l/hour must be 2 ±1 l/h. A
fl
ow value
below 0.5 or greater than 3.5 l/h generates a general fault alarm
displaying a Flow fault.
O2
Measured oxygen percentage
Return to Main Screen 3. The display returns to the Main Screen
automatically if the touch screen is not used for 2 minutes.
Adjust
Displays the adjustment screen to calibrate the analyzer.
Maintenance
Displays the analyzer's parameter setting screen (access code,
outputs, alarm limits, time-stamp, RS485,
fl
ow corrections).