XZR400 Series User’s Manual
Michell Instruments
15
INSTALLATION
2.8.7 Sample
Path
The MSRS sensor is placed inside an oven in which the gases to be analyzed are
circulated. The oven consists of a gas inlet head and a outlet plate. 3 Viton O-rings
ensure the sealing of this device (2 for the inlet and 1 for the outlet).
The internal sample path consists of:
• 2
fl
ow control valves: sensor
fl
ow and by-pass
fl
ow
• 1 sealing head
• 1 outlet plate
• 1 oven tube
• 1
electronic
fl
ow meter
• 1 MSRS sensor
• 3 Swagelok 6mm stainless steel bulkhead unions (1 gas inlet and 2
gas outlets) - on the rear panel
XZR400A1
Sensor flow adjustment valve
O
2
sensor
By-pass flow adjustment valve
Barometric pressure sensor
Ɨ
” bulkhead union
Ɨ
” bulkhead union
Ɨ
” bulkhead union
Sampling gas inlet
Electronic flowmeter 0-12 l/h
Sampling gas outlet
Sampling gas outlet
Figure 16
Sample Path XZR400A1