XZR400 Series User’s Manual
Michell Instruments
75
APPENDIX C
C.3
RS232 Command List
Mnemonic
ASCII Value
_OXY
O
2
concentration
XX.XX
or X.XXE±XX
_TEM
Oven temperature
XXX.XX
_UMV
MSRS voltage
XXX.XX
_AMB
Ambient temperature
XX.XX
_ALR
K1, K2, K3 Relay status
K1 +K2 +K3 = from 0 to 7
(1) (2) (4)
X
_CAL
Sets self calibration after 10 minutes bleed
RECEIPT
PURGE?
_FIN
Sets a 5 minute bleed in case of failure in calibration
RECEIPT
DEFAULT
_ACQ
Acknowledges the alarms
RECEIPT
_ETA
Display of calibration gas value
XX.XX
E_ETA_X.XX
Set of calibration gas value
X.XX
_STP
Display of oven temperature set point
XXX.XX
E_STP_XXX.XX
Set of oven temperature
XXX.XX
_AL1
Display of
fi
rst alarm level
XXX.XX
E_AL1
Set of
fi
rst alarm level
XX.XX
_AL2
Display of second alarm level
XXX.XX
E_AL2
Set of second alarm level
XX.XX
_NET
Starts self cleaning
RECEIPT
_YYY
Unknown entry
ERROR
_TCA
Measured oven tempe coef. due to
adjustment
XX.XX
_BRK
Allows for: Ending the bleed during calibration
Ending self cleaning and start cooling down
XX.XX
_PAB
Value of atmospheric pressure
XXXX.XX
_ indication corresponds to the space bar
indication corresponds to the return key
Table 2
RS232 Commands