XZR400 Series User’s Manual
Michell Instruments
61
CALIBRATION
Managing The Alarms During The Calibration Sequence
The alarms can be active or inactive during the calibration sequence.
When
NO
is on in the
Active during ADJ
section on screen page 5-3 the alarm contacts
are not active during the calibration sequence.
When
YES
is on in the
Active during ADJ
section on screen page 5-3 the alarm contacts
will be activated during the calibration sequence and will respond according to the
con
fi
guration set by the user.
De-Activation of Alarms
The alarms may be de-activated during installation and set-up. If the alarms are de-
activated and the system detects a fault and/or an alarm threshold is passed, the alarm
relays are not activated. In this case a message describing the fault will appear at the
bottom of the default screen page 3.
When
NO
is on in the
Inhibit Alarms
section the alarms are operative.
When
YES
is on in the
Inhibit Alarms
section the alarms are de-activated.