XZR400 Series User’s Manual
Michell Instruments
43
OPERATION
3.6.2 Alarms
The Main Tab
The Alarms Screen sets the general behaviour for alarms 1 and 2.
•
Touch the area to select it and change it.
• Click
OK
to con
fi
rm or
Cancel
to discard the changes.
Alarm 1
Alarms
5.3
No
Enable when adj
Disabled
No
Alarm 2
Main
Figure 38
Alarm Screen (Main)
Area
Function
Enable when adj
Sets the behaviour for alarms 1 and 2 exclusively during auto adjust
operations or manual.
•
No
: exceeding the alarm threshold will not trigger the
relay for alarms 1 and 2 during adjustment operations, or
when alarm set off is con
fi
gured by the user.
•
Yes
: exceeding the alarm threshold will trigger the relay
for alarms 1 and 2 during adjustment operations, or when
alarm set off is con
fi
gured by the user.
Disabled
Sets the current behaviour for alarms 1 and 2 and that of the General
fault exclusively in normal mode of operation (excluding adjustment).
•
No
: the alarms are operational; the corresponding alarm
will be activated when an alarm is triggered and/or a fault
occurs. This is the standard choice for the analyzer in
normal operation.
•
Yes
: no alarms will be activated regardless of the alarm and/
or fault present. Use this option during the commissioning
of the analyzer, especially before the wiring the alarms.
The blinking
Alarms Disabled
message will be displayed
on the Measurement Display Screen (3).
NOTE: Operating with alarms disabled must only be done
temporarily, particularly during adjustments.
Alarm 1 tab
Touch the area to display the alarm 1 threshold change screen.
Alarm 2 tab
Touch the area to display the alarm 2 threshold change screen.
Returns to the Main Menu Screen.