XZR400 Series User’s Manual
48
97472 Issue 3, August 2018
OPERATION
Process Tab
Main
Auto adjustment
5.4.1
Adjustment time
Control time
Purge time
Hold 4-20 mA
Alarms
Ctrl Adj Pressure
20 min
25 min
30 min
No
No
No
Level
Process
Area
Function
Adjustment
time
Duration in minutes of the adjustment sequence.
Control time
Duration in minutes of the control sequence.
Purge time
Duration in minutes of purging.
Hold 4-20 mA
Concerns the 4-20 mA analyzer outputs
•
No
: 4-20 mA outputs will display the oxygen concentration
during the adjusting sequence start until the purge end.
•
Yes
: 4-20 mA outputs are locked to the value measured
before the adjustment sequence start until the end purge.
Alarms
Concerns the analyzer general fault alarm as well as Alarms 1 and 2
•
No
: alarms will remain operational during the adjust
sequence start until the end of the purge.
•
Yes
: alarms will be disabled.
Ctrl Adj
Pressure
Gas adjustment and control gas bottle pressure checking
•
No
: any pressure control on bottles of gas adjustment or
gas control will be carried out. In case of fault in at least one
of these bottles, the conformity of the adjustment or control
cannot be guaranteed.
•
Yes
: the adjustment sequence or control will stop if any
pressure fault is detected. The
No adjusting
message
or
Control not possible
will be displayed. The previous
adjustment will be retained.
Returns to the Main Menu Screen.