MI 2892 Power Master
Operating the instrument
63
TREND
Switches to TREND view (available only during recording).
TABLE
Switches to TABLE view (available only during recording).
Moves cursor and select time interval (IP) for observation.
Returns to the “MEASUREMENTS” submenu.
3.13.3
Table
During active recording TABLE view is available (see section 3.14 for instructions how to start recording),
by cycling function key F4 (METER –TREND - TABLE). Signalling events can be here observed as required
by standard IEC 61000-4-30. For each signalling event instrument capture waveform which can be
observed in PowerView.
Figure 3.45: Signalling table screen
Table 3.52: Instrument screen symbols and abbreviations
No
Signalling event number
L
Phases on which signalling event occurred
F
Flag indication
0 – none of intervals are flagged
1 – at least one of intervals inside recorded signalling is
flagged
Sig
Frequency on which signalling occurred, defined as “Sign. 1”
frequency (f1) and “Sign. 2” frequency (f2) in SIGNALLING SETUP
menu. See 3.21.4 for details.
START
Time when observed Signalling voltage crosses threshold boundary.
MAX
Maximal voltage level recorder captured during signalling events
Level
Threshold level in % of nominal voltage Un, defined in SIGNALLING
SETUP menu. See 3.21.4 for details.
Duration
Duration of captured waveform, defined in SIGNALLING SETUP
menu. See 3.21.4 for details.
f1
1
st
observed signalling frequency.
f2
2
nd
observed signalling frequency.
F4