MI 2892 Power Master
Operating the instrument
64
Table 3.53: Keys in Signalling (TABLE) screen
METER
Switches to METER view.
TREND
Switches to TREND view (available only during recording).
TABLE
Switches to TABLE view (available only during recording).
Moves cursor through signalling table.
Returns to the “MEASUREMENTS” submenu.
3.14
General Recorder
Power Master has ability to record measured data in the background. By entering GENERAL RECORDER
option from RECORDERS submenu, recorder parameters can be customized in order to meet criteria
about interval, start time and duration for the recording campaign. General recorder setup screen is
shown below:
Figure 3.46: General recorder setup screen
Description of General recorder settings is given in the following table:
Table 3.54: General recorder settings description and screen symbols
General recorder is active, waiting for start condition to be met.
After start conditions are met (defined start time), instrument will
capture waveform snapshot and start (activate) General recorder.
General recorder is active, recording in progress
Note: Recorder will run until one of the following end conditions is
met:
STOP key was pressed by user
Given Duration criteria was met
Maximal record length was reached
SD CARD is full
Note: If recorder start time is not explicitly given, recorder start
depends on Real Time clock multiple of interval. For example:
recorder is activated at 12:12 with 5 minute interval. Recorder will
actually start at 12:15.
F4