3.8.4.6 Differential Relay Instantaneous Test Results
This screen displays the results upon test completion. No data may be entered in this screen.
Figure 56 Differential Instantaneous Test Screen
The Timing, Target & Seal-In, and Pickup test screens may be entered to change test settings.
Each current channel must be selected by touching I1, I2, and I3. However, there must be a
corresponding V/I Generator.
The bottom of the screen has common buttons.
File:
This selection will take the user to the File
Manager.
Batt Sim Off / Batt Sim On:
This selection will turn the Battery Simulator On
and Off.
Test Menu:
This selection will return the controller to the
Test Menu Screen.
Main Menu:
This selection will return the controller to the
Main Menu:
Start / Stop:
This selection will apply the set quantities
displayed and begin the test. After Start is
pressed this button will turn red and display
Stop. Pressing Stop at any time will stop the
test and turn the outputs off.