Condition
Sets the parameters under which a true or false contact
is determined. The selections of condition are as
follows:
NO
Normally Open dry contact.
NC
Normally Closed dry contact.
VA
Voltage Applied wet contact.
VR
Voltage Removed wet contact.
N/A
Not
Applicable.
Latch Mode:
Sets the parameter under which the timer will either stay
on after a contact state becomes true or will return to
previous state after the contact changes state.
Latch On: Will latch the timer in that condition after the
timer has reached a true condition.
Latch Off: Will allow the timer to return to original
condition even after the contact has reached a true
condition.
Max Operate Time
Determines the maximum timer the unit will operate
before the outputs are turned off. This function is only
applicable when the timer is operating.
Debounce Time
Determines how long a contact must be in a true state
before the condition is determined to be true.
Turn Currents Off (Stop Contact Only) When selected will turn all output currents off when the
stop contact becomes true.
Turn Voltages Off (Stop Contact Only) When selected will turn all output voltages off when the
stop contact becomes true.
Horn On / Horn Off
This selection will turn the horn on when any contact is
determined to be true
Seconds / Cycles
This selection will toggle the timer between measuring in
seconds or cycles.
Complex Timer
This selection will display the Complex Timer Setup
menu.
Timer Reset
This selection will reset the timer to zero.
Back.
This selection will return the user to the screen
previously displayed.