Amplifier EXT Inputs
– There can be up to four sets of BNC connectors labeled V1/I5, I1; V2/I6,
I2; V3/I7, I3 and V4/I8, I4 on the back panel depending on the number of amplifier modules
installed. These connectors are used to amplify an external analog signal using the MPRT
amplifiers. Application of ± 10 Volts Peak will provide Full Scale output from the selected output.
Figure 3 MPRT Rear Panel, Analog Input Terminals
CAUTION
: DO NOT APPLY MORE THAN ± 10 VOLTS PEAK TO THE EXT INPUT
TERMINALS. APPLICATIION OF MORE THAN 10 VOLTS PEAK MAY DAMAGE THE
AMPLIFIER.
1.1.3
Touch View Interface
The Touch View Interface is used in manual operation of the unit. The color LCD touch-panel
display and control knob are the operator interface for the unit.
Figure 4 Touch-View Interface
1.
LCD Color Display
– this is an 8.5 inch touch panel display.
2.
Control Knob
– this knob will adjust values once the cursor is in the box location of the
value to be changed.