62
7.1
Drain the Control Module
Tools and Supplies Required:
Arctic Sun Drain Bag
1. Turn the Control Module off.
Caution
: Draining the system with power on may damage the
chiller.
2. Attach the two fittings on the bag to the two drain ports on the
back of the control module. The device will passively drain all
tubing, reservoirs and pumps within the system. This is an
adequate drain for maintenance procedures.
Drain Bag Attached
Figure 7.5
7.2 Front
Grill
1. Remove the front grill by first pulling the bottom out by reaching
under the center of the bottom edge.
2. Using both hands at the top sides of the panel, pull it straight
out and away from the control module.
3. To replace, align the studs with the holes and push all four
corners.
Front Grill Removal
Figure 7.6
7.3 Rear
Panel
Tools and Supplies Required:
5/32” hex key
1. Turn off the power and unplug the power cord.
2. Detach all cables and lines from the rear panel: fluid delivery
line, fill line, patient temperature cables, remote display cable
and strain relief.
3. Remove the four mounting screws at the corners with the hex
key from the rear panel.
4. Pull the panel off horizontally.
Rear Panel Removal
Figure 7.7