Table 6.4
Alarm and Alert Troubleshooting
Alarm/
Alert
Message Displayed
Alarm/Alert
Description
System Status
Troubleshooting
86 Non-recoverable
system error
Refer to Operator
Manual.
Alarm
Power supply
voltage out of range.
Changes to Stop
Mode. Audio alarm
half second tone
every 10 seconds.
Alarm/Alert indicator
flashing.
Turn the Control Module off. Wait 30 seconds and
turn power back on. If the error persists, replace
the Power Card and attempt a power-up. If the
error persists, call Medivance Service. Calibrate
upon completion.
98.x Non-recoverable
system error
Refer to Operator
Manual.
Alarm
Processor fault
Changes to Stop
Mode. Audio alarm
half second tone
every 10 seconds.
Alarm/Alert indicator
flashing.
Replace the Processor and retest. Calibrate upon
completion.
99 Non-recoverable
system error
Refer to Operator
Manual.
Alarm
Processor Fault
Changes to Stop
Mode. Audio alarm
half second tone
every 10 seconds.
Alarm/Alert indicator
flashing.
Replace the Processor Card. Calibrate upon
completion.