45
Table 6.4
Alarm and Alert Troubleshooting
Alarm/
Alert
Message Displayed
Alarm/Alert
Description
System Status
Troubleshooting
08 System
warming
and
Patient Temperature
1 high.
Alarm
Patient Temperature
1 reading is above
38.
◦
C (100.4°F) and
water temperature is
above 38.
◦
C
(100.4°F) and
system is unable to
cool in Automatic
Mode
Changes to Stop
Mode. Audio alarm
half second tone
every 2 minutes.
Alarm indicator
flashing
Operate the system in Manual Mode with a setpoint
below 38°C until the water temperature is less than
38°C. Press Automatic and retest. If the error
persists, see General Troubleshooting - Will Not
Cool.
09 Patient
Temperature
1 above user defined
limit.
Alert
Patient Temperature
1 reading is equal or
greater than the
user-defined value
established in the
custom settings
menu.
Current Mode
Continues. Audio
alarm half second
tone every 2
minutes. Alarm/Alert
indicator flashing
Adjust Patient High Temperature Alert in the
custom menu if desired.
10
System cooling and
Patient Temperature
1 low.
Alarm
Patient Temperature
1 reading is below
32.0
◦
C (89.6°F) and
water temperature is
below 32.0
◦
C
(89.6°F) and the
system is unable to
heat in Automatic
Mode
Changes to Stop
Mode. Audio alarm
half second tone
every 10 seconds.
Alarm/Alert indicator
flashing
Operate the system in Manual Mode with a setpoint
above 32°C until the water temperature is greater
than 32°C. Press Automatic and retest. If the
error persists, see General Troubleshooting - Will
Not Heat.
11 Patient
Temperature
1 below user defined
limit
Alert
Patient Temperature
1 reading is equal or
less than the user-
defined value
established in the
custom settings
menu.
Current Mode
continues. Audio
alarm half second
tone every 2
minutes. Alarm/Alert
indicator flashing.
Adjust Patient High Temperature Alert in the
custom menu if desired.
12 Patient
and
water
temperature high.
Check water
temperature.
Alert
Patient Temperature
1 reading is above
38.0
◦
C (100.4°F)
and water
temperature is above
38.0
◦
C (100.4°F)
and the system is
unable to cool in
Manual Mode
Current Mode
continues. Audio
alarm half second
tone every 2
minutes. Alarm/Alert
indicator flashing.
Operate the system in Manual Mode with a setpoint
below 38°C until the water temperature is less than
38°C. If the error persists, see General
Troubleshooting - Will Not Cool.