17
CHAPTER 4 MAINTENANCE
4.1
Recommended Maintenance Schedule
Procedure Interval
Clean External Surfaces
As required
Adjust Remote Display Clamp
As required
Inspect connectors and cables
3 months
Clean the Condenser
3 months
Replace Filter
3 months
Replace Fluid
3 months
Replace Internal Filters
12 months
Calibration and Operational Check
Every 2000 hours or 250 uses, which ever is
less, as indicated by system display
4.2
Clean External Surfaces
Cleaning should include the exterior of the Control Module and the power cord. Use a soft cloth and
mild detergent or disinfectant according to hospital protocol. The display screen can be wiped with a
clean, lint-free, moist cloth. Fluid Delivery Lines and Temperature Cables should be cleansed with a
hospital approved disinfectant.