33
The history logs are interpreted as follows:
Table 6.2
Log Definitions
Field
Definition
Units or Range of Values
C
Error Code
See Table 6.4
M
Control Module Operating Mode
0 = Initialization, 1= Stop, 2 =
Automatic, 3 = Manual, 4 = Purge,
5 = Fill
P
Patient Temperature 1
°
C
OM
Outlet Temperature Monitor (T1)
°
C
O
Outlet Temperature Control (T2)
°
C
I
Inlet Temperature (T3)
°
C
F Flow
Rate
ml/min
P Inlet
Pressure
0 to -10 psi
L Water
Level
1 = Empty, 5 = Full
P
Circulating Pump PWM
0= Stop to 235 = Full
C Mixing
Pump
PWM
0 = Stop to 150 = Full
H Heater
PWM
0 = Off to 32 = Full Power
V
Vent Valve State
1=Open, 0 = Closed
B
Bypass Valve State
1=Open, 0 = Closed
Hr System
Hours
Hours