32
6.1.1 Diagnostic Home Screens
Toggle between these two diagnostic displays by pressing the Home key.
Diagnostic Home Page #1
Figure 6.1
Diagnostic Home Page #2
Figure 6.2
The display fields are:
Table 6.1
Display Field Definitions
Field
Definition
Units or Range of Values
PT1
Patient Temperature #1
°
C
PT2
Patient Temperature #2
°
C
FL Water
Flow
Rate
ml/min
T1
Monitor Water Temperature:
°
C
T2
Control Water Temperature:
°
C
T3
Return from Pad Water Temperature
°
C
T4
Chiller Loop Water Temperature
°
C
IP
Inlet Water Pressure
0 to -10 psi
L
Reservoir Level
1 = Empty, 5 = Full
VV
Vent Valve
1=Open, 0 = Closed
BV
Bypass Valve
1=Open, 0 = Closed
PWM P Circulating Pump Speed
0= Stop to 235 = Full
PWM M Mixing Pump Speed
0 = Stop to 150 = Full
PWM H Heater Power
0 = Off to 32 = Full Power
6.1.2 Error Log and Operational Log
The Model 2000 maintains two history logs. The logs may be accessed while in Stop mode, once
Diagnostic Mode is activated, by pressing the Down Arrow key. The Error Log screens are labeled
UD2 and the Operational Log screens are labeled UD3. The Operational Log contains a history of
the last 16 alerts and recoverable alarms. The Error Log contains a history of the last 8 non-
recoverable alarms. Note that power failures do not generate a logged alarm. Each screen provides
information on the state of the Control Module when the event occurred including the current system
hours.
Error Log Display
Figure 6.3