55
Table 6.4
Alarm and Alert Troubleshooting
Alarm/
Alert
Message Displayed
Alarm/Alert
Description
System Status
Troubleshooting
79 Non-recoverable
system error
Refer to Operator
Manual.
Alarm
T1 and T2 differ by
more than 1.0
degree C
Changes to Stop
Mode. Audio alarm
half second tone
every 10 seconds.
Alarm/Alert indicator
flashing.
Turn the Control Module off. Wait 30 seconds and
turn power back on. If the error persists, call
Medivance Service.
80 Non-recoverable
system error
Refer to Operator
Manual.
Alarm
The Control
Processor has failed
to detect a simulated
water temperature
fault.
Changes to Stop
Mode. Audio alarm
half second tone
every 10 seconds.
Alarm/Alert indicator
flashing.
Replace the I/O Card and attempt to power-up. If
the error persists, call Medivance Service.
Calibrate upon completion.
81 Non-recoverable
system error
Refer to Operator
Manual.
Alarm
The Monitor
Processor has failed
to detect a simulated
water temperature
fault.
Changes to Stop
Mode. Audio alarm
half second tone
every 10 seconds.
Alarm/Alert indicator
flashing.
Replace the I/O Card and attempt to power-up. If
the error persists, call Medivance Service.
Calibrate upon completion.
82 Non-recoverable
system error
Refer to Operator
Manual.
Alarm
A/D converter fault
Changes to Stop
Mode. Audio alarm
half second tone
every 10 seconds.
Alarm/Alert indicator
flashing.
Turn the Control Module off. Wait 30 seconds and
turn power back on. If the error persists, replace
the Processor Card. If unsuccessful, call
Medivance Service. Calibrate upon completion.
83 Non-recoverable
system error
Refer to Operator
Manual.
Alarm
The Monitor
Processor does not
communicate.
Changes to Stop
Mode. Audio alarm
half second tone
every 10 seconds.
Alarm/Alert indicator
flashing.
Turn the Control Module off. Wait 30 seconds and
turn power back on. If the error persists, replace
the Processor Card. If unsuccessful, call
Medivance Service. Calibrate upon completion.
84 Non-recoverable
system error
Refer to Operator
Manual.
Alarm
The Control
Processor does not
communicate.
Changes to Stop
Mode. Audio alarm
half second tone
every 10 seconds.
Alarm/Alert indicator
flashing.
Turn the Control Module off. Wait 30 seconds and
turn power back on. If the error persists, replace
the Processor Card. If unsuccessful, call
Medivance Service. Calibrate upon completion.
85 Non-recoverable
system error
Refer to Operator
Manual.
Alarm
The model of the
Control Module
could not be verified.
Changes to Stop
Mode. Audio alarm
half second tone
every 10 seconds.
Alarm/Alert indicator
flashing.
Call Medivance Service.