15
3.2
Power Up System
1. Insert the power cord into the power inlet. Using the clips and wing
nuts from the Strain Relief Kit, anchor the cord to the back panel.
Plug the power cord into the wall outlet. Locate the switch on the
rear panel and turn the power on.
2. After the software initializes an alarm will sound and the display
will read: “Alarm 5: Water level below minimum. Fill reservoir or
purge pads”.
3. Clear the alarm by pressing the alarm key.
4. The display reads:” Medivance Arctic Sun Temperature
Management System”.
3.3
Filling the Control Module
1. Add one vial of Arctic Sun Cleaning Solution to 4 liters of sterile or
distilled water. The solution is contained in the Arctic Sun
Maintenance Kit.
Caution:
Use only sterile or distilled water. The use of other fluids
will damage the system.
2. From Stop Mode, press the down arrow until the fill screen
appears: “Water Level Empty. Enter to Fill”. Press Enter to
proceed.
Note:
Pads cannot be attached to the Fluid Delivery Line during
the fill process.
3. Follow directions on the display screen which read:” Place the Fill
Tube in the container of water. Press Enter to fill”.
4. Press Enter again and the Model 2000 will automatically fill until
the reservoir is full.