22
Replace the Mixing Pump Filter
1
Remove the front grill by first pulling the bottom out by
reaching under the center of the bottom edge.
2
Using both hands at the top sides of the panel, pull it straight
out and away from the control module.
3
Remove the sheet metal access panel under the grill by
unscrewing the four nuts with the nut driver.
4
Locate the latch below the tube fitting on the back of the
Mixing Pump. Push the latch up to release the tube fitting
from the pump. Pull fitting away from you to disengage the
fitting from the pump.
5
Remove the filter from the fitting and replace with a new one.
Insert the open end of the new filter into the fitting leaving the
filter extended from the fitting as shown in Figure 4.10 above.
6
Insert the tube fitting back into the pump. Push the lever
down to lock the fitting.
Note:
Some water will drain from the tube after removal of the
fitting.
Access Panel Removal
Figure 4.11
Mixing Pump Filter Location
Figure 4.12
4.9
Calibration and Safety Check
See Chapter 5 for the complete procedure for calibration and safety check.