53
Table 6.4
Alarm and Alert Troubleshooting
Alarm/
Alert
Message Displayed
Alarm/Alert
Description
System Status
Troubleshooting
60 Non-recoverable
system error
Refer to Operator
Manual.
Alarm
Control and Monitor
start up
synchronization fault
Changes to Stop
Mode. Audio alarm
half second tone
every 10 seconds.
Alarm/Alert indicator
flashing.
Turn the Control Module off. Wait 30 seconds and
turn power back on. If the error persists, replace
the Processor Card. If unsuccessful, call
Medivance Service. Calibrate upon completion.
61 Non-recoverable
system error
Refer to Operator
Manual.
Alarm
Control Processor
parameter memory
fault
Changes to Stop
Mode. Audio alarm
half second tone
every 10 seconds.
Alarm/Alert indicator
flashing.
Turn the Control Module off. Wait 30 seconds and
turn power back on. If the error persists, replace
the Processor Card. If unsuccessful, call
Medivance Service. Calibrate upon completion.
62 Non-recoverable
system error
Refer to Operator
Manual.
Alarm
Monitor Processor
parameter memory
fault
Changes to Stop
Mode. Audio alarm
half second tone
every 10 seconds.
Alarm/Alert indicator
flashing.
Turn the Control Module off. Wait 30 seconds and
turn power back on. If the error persists, replace
the Processor Card. If unsuccessful, call
Medivance Service. Calibrate upon completion.
64 Non-recoverable
system error
Refer to Operator
Manual.
Alarm
Unable to enable
pump power
(Control)
Changes to Stop
Mode. Audio alarm
half second tone
every 10 seconds.
Alarm/Alert indicator
flashing.
Replace the Power Card and attempt to power-up.
If the error persists, call Medivance Service.
65 Non-recoverable
system error
Refer to Operator
Manual.
Alarm
Unable to enable
pump power
(Monitor)
Changes to Stop
Mode. Audio alarm
half second tone
every 10 seconds.
Alarm/Alert indicator
flashing.
Replace the Power Card and attempt to power-up.
If the error persists, call Medivance Service.
66 Non-recoverable
system error
Refer to Operator
Manual.
Alarm
Unable to disable
pump power
(Control)
Changes to Stop
Mode. Audio alarm
half second tone
every 10 seconds.
Alarm/Alert indicator
flashing.
Replace the Power Card and attempt to power-up.
If the error persists, call Medivance Service.
67 Non-recoverable
system error
Refer to Operator
Manual.
Alarm
Unable to disable
pump power
(Monitor)
Changes to Stop
Mode. Audio alarm
half second tone
every 10 seconds.
Alarm/Alert indicator
flashing.
Replace the Power Card and attempt to power-up.
If the error persists, call Medivance Service.
70 Non-recoverable
system error
Refer to Operator
Manual.
Alarm
Mains signal is
missing or
inconsistent
Changes to Stop
Mode. Audio alarm
half second tone
every 10 seconds.
Alarm/Alert indicator
flashing.
Replace the Power Card and attempt to power-up.
If the error persists, call Medivance Service.