52
Table 6.4
Alarm and Alert Troubleshooting
Alarm/
Alert
Message Displayed
Alarm/Alert
Description
System Status
Troubleshooting
48 Patient
temperature
out invalid.
Alert
Patient Temperature
Output (Echo)
calibration data in
non-volatile memory
invalid.
Current Mode
Continues. Audio
alarm half second
tone every 2
minutes. Alarm/Alert
indicator flashing.
Turn the Control Module off. Wait 30 seconds and
turn power back on. If the error persists, replace
the Processor Card. If unsuccessful, call
Medivance Service. Calibrate upon completion.
50 Patient
Temperature
1 erratic. Check
temperature probe.
Alert
Patient temperature
discontinuity.
Significant change in
patient temperature
reading for more
than 8 minutes.
Current Mode
Continues. Audio
alarm half second
tone every 2
minutes. Alarm/Alert
indicator flashing
Use the CTU to verify the accuracy of the Control
Module. Check connection of patient temperature
probe and cable. If wet, replace cable. Check
temperature cable connection at the rear of the
Control Module. Flex the cable to ensure that there
is not an intermittent fault. Test the input with a
cable and probe proven good on another Control
Module. If the Control Module proves inaccurate,
replace the Isolation Card and retest. If the error
persists, call Medivance Service. Recalibrate upon
completion.
51 Patient
Temperature
1 below control
range. Check
temperature probe.
Alert
Patient temperature
less than 33
°
C when
Automatic mode
entered.
Current Mode
Continues. Audio
alarm half second
tone every 2
minutes. Alarm/Alert
indicator flashing
Use the CTU to verify the accuracy of the Control
Module. Check connection of patient temperature
probe and cable. If wet, replace cable. Check
temperature cable connection at the rear of the
Control Module. Flex the cable to ensure that there
is not an intermittent fault. Test the input with a
cable and probe proven good on another Control
Module. If the Control Module proves inaccurate,
replace the Isolation Card and retest. If the error
persists, call Medivance Service. Recalibrate upon
completion.
52
Extended period of
cold water delivery.
Assess Patient
Alert
Circulating water
temperature has
been below
10°C/50°F for 8 of
the previous 10
hours.
Remains in current
mode. Audio
alarms.
Assess patient’s skin. Check to determine if water
circulating through the pads is at least 2.3 liters per
minute. Determine whether the temperature probe
is placed properly. Check to make sure the
recommended number of pads is in use. Determine
whether the patient requires additional sedation for
shivering. Assess whether environmental factors,
such as room temperature and hot lamps are
interfering with the cooling process.
53 Prolonged
cold
exposure
Alarm
The Arctic Sun has
issued 11 alerts for
extended periods of
cold water delivery.
Changes to Stop
Mode
Assess patient’s skin. Assess whether the minimum
water temperature should be increased or whether
the temperature management therapy should be
discontinued.