M-9500-3.0.6-0 INTEGRA 9500-9505 Operations Manual – English
Approved
: Spike Reed
Effective:
15 Nov 2010
Supersedes:
28 Jan 2010
LVS®
CONFIDENTIAL
All contents copyright © LVS® 2010 Page 62 of 82
Note:
If “Modulation” is selected, the modulation for each dark/white
transition is displayed. The lowest modulation for that scan line is
displayed in red.
Decodability (width of each element)
Defects (inflection to element reflectance)
Full screen waveform
Traditional bar growth and shrinkage
The background of the
SRP View Tab
screen represents the bars and spaces of
the actual barcode symbol. The global threshold is also displayed.
The PgUp and Pgdn arrows on the keyboard allow you to inspect every scan
line. Each scan line is approximately 0,05 mm (.002”) apart.
Full Screen Waveform
The Full Screen Waveform feature allows you to view the entire Scan
Reflectance Profile at one time (see Figure below).
Note:
If a quiet zone failure occurs, click the viewing box to locate the failure.
The SRP window will show the actual quiet zone and required quiet zone (see
Figure below).