Calibration Tab
The Calibration tab (see Figure below) is where the calibration of the system
occurs.
One of the following two Calibrated Conformance Standard Test Cards were
supplied with your INTEGRA 9500/9505. Note that the Test Card you receive
depends on your field of view.
GS1-128
EAN/UPC
Symbol
Verifiers
The Calibrated Conformance Standard Test Cards are used to certify that the
system is calibrated according to the ISO/IEC 15416:2000(E) and is traceable to
NIST standards. It is specially printed to have certain reflective values.
Note that
the INTEGRA 9500/9505 recognizes only these cards for calibration.
The upper left corner of the
Calibration Tab
screen shows the date/time stamp
that is placed on an internal log designed to archive all calibration activities. It
lists a date/time stamp along with the person who performed the calibration. This
report can be viewed or printed. The log is a “read-only” file and cannot be
altered or deleted. To view the calibration log, go to the
Archive Tab
screen.