M-9500-3.0.6-0 INTEGRA 9500-9505 Operations Manual – English
Approved
: Spike Reed
Effective:
15 Nov 2010
Supersedes:
28 Jan 2010
LVS®
CONFIDENTIAL
All contents copyright © LVS® 2010 Page 53 of 82
4. OCR
To check the human readable characters, right-click on the mouse and draw
a box around the human readable characters; the box appears in green. The
software reports a “match” if it matches the decoded barcode information.
The system supports OCR-A, OCR-B, Times New Roman, Arial, Courier, and
most San-serif fonts. Upper case only. Special characters are not supported.
The human readable of a barcode label can be checked most of the time
provided the characters do not touch each other. The characters must also be
large enough to read and in a straight line. The system is limited to reading up
to 24 characters.
5. Modulation
To view a modulation error, click the
Modulation
option. It is recommended
to activate the Zoom button along with the Modulation button so that the
image is large enough to see the details.
When viewing Modulation errors, the software may use a “yellow” mark. A
yellow mark indicates that the Modulation measurement for that cell was less
Thumbnail view in color
according to the legend.
Color legend