Using the Neon
™
Transfection System,
Continued
Electroporation
Protocol,
Continued
17.
Repeat Steps 7–16 for the remaining samples.
Be sure to change the Neon
™
Tips after using it twice and Neon
™
Tubes
after 10 usages. Use a new Neon
™
Tip and Neon
™
Tube for each new
plasmid DNA sample.
18.
Gently rock the plate to assure even distribution of the cells. Incubate the
plate at 37°C in a humidified CO
2
incubator.
19.
If you are not using the Neon
™
device, turn the power switch on the rear
to
OFF
.
20.
Assay samples to determine the transfection efficiency (e.g., fluorescence
microscopy or functional assay) or gene knockdown (for siRNA).
Optimization
Based on your initial results, you may need to optimize the electroporation
parameters for your cell type. See page 22 for using the 18-well or
preprogrammed 24-well optimization protocol on the Neon
™
device.
Cleaning and
Maintenance
Clean the surface of the Neon
™
device and Neon
™
Pipette Station with a damp
cloth.
Do not
use harsh detergents or organic solvents to clean the unit. The
Neon
™
Pipette is permanently calibrated at the manufacturer and does not
require any further calibration.
Important!
Avoid spilling any liquid inside of the Neon
™
Pipette Station to
prevent any build up of rust on the ball plunger in the pipette station.
In case you accidentally spill any liquid (e.g., buffer, water, coffee) inside the
Neon
™
Pipette Station, disconnect the station from the main device and wipe
the station using dry laboratory paper. Invert and allow the station to
completely dry for 24 hours at room temperature.
Do not use the oven to dry
the Neon
™
Pipette Station.
If the station does not work after drying, contact
Technical Support (page 39).
For any other repairs and service, contact Technical Support (page 39).
Do not
perform any repairs or service on the
Neon
™
device yourself as it is a high
voltage hazard and to avoid any damage to the unit or voiding your warranty.
21