Using the Neon
™
Transfection System,
Continued
Electroporation
Protocol,
Continued
11.
Insert the Neon
™
Pipette with the sample vertically into the Neon
™
Tube
placed in the Neon
™
Pipette Station until you hear a click sound. Ensure that
the pipette projection is inserted into the groove of the pipette station.
Note:
Ensure the metal head of the Neon
™
Pipette is tightly connected to the ball
plunger inside of the Neon
™
Pipette Station and to the Neon
™
Tube (see figure on the
left for the correct position).
O
ⅹ
O
ⅹ
12.
Ensure that you have selected the appropriate electroporation protocol and
press
Start
on the touchscreen.
13.
The Neon
™
device automatically checks for the proper insertion of the Neon
™
Tube and Neon
™
Pipette before delivering the electric pulse.
Note:
Monitor the Neon
™
Tip during electroporation to see if there is any arcing
(sparks) that is caused by the presence of bubbles in the tip. Arcing results in low
transfection efficiency and cell viability.
14.
After delivering the electric pulse,
Complete
is displayed on the touchscreen
to indicate that electroporation is complete.
15.
Slowly remove the Neon
™
Pipette from the Neon
™
Pipette Station and
immediately transfer the samples from the Neon
™
Tip by pressing the push-
button on the pipette to the first stop into the prepared culture plate
containing prewarmed medium.
Note:
We strongly recommend loading electroporated cells into growth medium
without antibiotics
that can greatly reduce the viability of your cells after transfection.
16.
To discard the Neon
™
Tip, press push-button to the second stop into an
appropriate biological hazardous waste container.
Continued on next page
20