Description of Parts,
Continued
Neon
™
Pipette
Station
The Neon
™
Pipette Station is a unique component of the Neon
™
Transfection
system. It holds a Neon
™
Pipette during electroporation procedures. The
Neon
™
Pipette Station is equipped with many safety sensors and protection
mechanisms that protects the user from any exposures to an electrical shock.
The Neon
™
Pipette Station is connected to the Neon
™
device using the high
voltage and sensor connector (see page 6 for details).
The Neon
™
Pipette Station also holds the Neon
™
Tube which has an electrode
near the bottom that transfers the electric field from the electrode inside the
Neon
™
Tip.
Area to insert
Neon
™
Tube
Neon
™
Pipette
Station
Connector
Neon
™
Tube
The Neon
™
Tube holds the Electrolytic Buffer during electroporation and is
inserted into the Neon
™
Pipette Station. The Neon
™
Pipette with the Neon
™
Tip is
then inserted into the Neon
™
Tube which has an electrode near the bottom that
transfers the electric field from the electrode inside the Neon
™
Tip. The Neon
™
Tubes are supplied with Neon
™
Kits as well as available separately (page 38).
To avoid contamination, we strongly recommend using the tubes for a
maximum of 10 times only.
We recommend changing tube and buffer when
switching to a different plasmid DNA/siRNA or cell type.
Tube Specifications:
Material: Polystyrene
Capacity: 2.5
–
4 mL
Buffer
Electrode
Continued on next page