Using the Neon
™
Transfection System,
Continued
Setting up the
Neon
™
Pipette
Station
1.
Ensure the Neon
™
Pipette Station is connected to the Neon
™
device (page 6).
2.
Fill the Neon
™
Tube with 3 mL of Electrolytic Buffer
(use Buffer E for 10 μL
Neon
™
Tip and Buffer E2 for 100 μL Neon
™
Tip)
.
Note:
Make sure that the electrode on the side of the tube is completely immersed in
buffer.
3.
Insert the Neon
™
Tube into the Neon
™
Pipette Station until you hear a click
sound.
Note:
Make sure that the side electrode of the Neon
™
tube is well connected to the
side ball plunger of the Neon
™
Pipette Station (see figure on the left below for correct
position).
O
Ball plunger
Side electrode
ⅹ
O
Ball plunger
Side electrode
ⅹ
4.
The station is ready for use. Proceed to preparing cells, next page.
15