on the first signal. For State Qualified triggers the amplitude of the first signal must remain in the
desired state until the trigger occurs. A qualified trigger can occur immediately after the validation,
or following a predetermined time delay or number of potential trigger events. The time delay or
trigger count is restarted with every validation.
Within Time
creates a time window within which a trigger can occur.
Wait Time
determines a delay from the start of the desired pattern. After the delay
(timeout) and while the pattern is present, a trigger can occur. The timing for the delay is restarted
when the selected pattern begins.
Events
determines a minimum number of events of the trigger source. An event is
generated when a trigger source meets its trigger conditions. On the selected event of the trigger
source and while the pattern is present, a trigger can occur. The count is initialized and started
whenever the selected pattern begins, and continues while the pattern remains. When the
selected count is reached, the trigger occurs.
Edge Qualified and Wait: Trigger after timeout. The broken upward-pointing arrows indicate potential triggers,
while the bold ones show where the actual triggers occur.
Summary of Contents for 6000 SERIES
Page 65: ...WR6K OM E Rev A ISSUED October 2003 63 Proper Orientation of Drive ...
Page 131: ...WR6K OM E Rev A ISSUED October 2003 129 Example ...
Page 141: ...WR6K OM E Rev A ISSUED October 2003 139 ...
Page 236: ...Operator s Manual 234 ISSUED October 2003 WR6K OM E Rev A Example 6 ...
Page 261: ...WR6K OM E Rev A ISSUED October 2003 259 ...
Page 286: ...Operator s Manual 284 ISSUED October 2003 WR6K OM E Rev A ...
Page 304: ...Operator s Manual 302 ISSUED October 2003 WR6K OM E Rev A ...
Page 307: ...WR6K OM E Rev A ISSUED October 2003 305 Convolve two signals ...
Page 348: ...Operator s Manual 346 ISSUED October 2003 WR6K OM E Rev A ...
Page 402: ...Operator s Manual 400 ISSUED October 2003 WR6K OM E Rev A ...