Operator's Manual
376
ISSUED: October 2003
WR6K-OM-E Rev A
Safe Operating Area Measurement – Steady State
Note:
The previous sections should be completed before you make these measurements.
Selecting
Safe Operating Area
in the "Functions" menu brings up the following display. The
voltage waveform (
V
) is displayed in the first grid and the deskewed current waveform (
I
) is
displayed in the second grid. In the XY plot, voltage points are plotted on the horizontal axis, while
current is plotted on the vertical. The delay between the voltage and current samples have been
removed by the deskew function.
Fine Adjust DC Level
fields for both voltage and current are provided to allow you to
compensate for any residual DC offset that the voltage and current probes may exhibit. Because
the magnitude of residual DC offset error is usually unknown, use of these adjustments requires
that you know something about the waveforms.
1.
Touch the Voltage (
V
) and Current (
I
) descriptor labels
in turn to open the setup dialogs for each.
2.
Press the
Fine
button, then turn the front panel
Adjust
knob to set the DC Offset Level.
3.
For a detailed explanation of how to adjust the DC Offset Level, see Fine DC Level
Adjustment.
Summary of Contents for 6000 SERIES
Page 65: ...WR6K OM E Rev A ISSUED October 2003 63 Proper Orientation of Drive ...
Page 131: ...WR6K OM E Rev A ISSUED October 2003 129 Example ...
Page 141: ...WR6K OM E Rev A ISSUED October 2003 139 ...
Page 236: ...Operator s Manual 234 ISSUED October 2003 WR6K OM E Rev A Example 6 ...
Page 261: ...WR6K OM E Rev A ISSUED October 2003 259 ...
Page 286: ...Operator s Manual 284 ISSUED October 2003 WR6K OM E Rev A ...
Page 304: ...Operator s Manual 302 ISSUED October 2003 WR6K OM E Rev A ...
Page 307: ...WR6K OM E Rev A ISSUED October 2003 305 Convolve two signals ...
Page 348: ...Operator s Manual 346 ISSUED October 2003 WR6K OM E Rev A ...
Page 402: ...Operator s Manual 400 ISSUED October 2003 WR6K OM E Rev A ...